ACE-828M
ACE-828M Image
This image is representative of the product and may not be the exact image of this product.
Model: ACE-828M
Manufacturer: IEI
MEDICAL AXT 280W Power Supply
Product Status: Discontinued - This product is no longer in production. This page is maintained for informational purposes only.

Stock Status: Not Available

 
 Input Table :
  • Input Table :

Model

Input Range

Input Frequency

ACE-828M

90 ~ 130V AC or 180 ~ 260V AC ( Auto-Switch )

47 ~ 63Hz

ACE-925M

90 ~ 264V AC ( Full-Range )

47 ~ 63Hz

  • Output Table : ( at 40oC )

Model

Rated Load

Max Output

+5V

+12V

-12V

-5V

-3.3V

+5Vsb

+5V

+12V

-12V

-5V

-3.3V

+5Vsb

ACE-828M

20A

10A

0.5A

0.5A

8A

0.72A

30A

15A

1A

1A

15A

1.2A

ACE-925M

25A

7A

0.5A

0.5A

30A

10A

1A

1A

  • Inrush Current :
    -->
    ACE-828M : Not exceed 40A@115V AC or 80A@230V AC, cold start at 25oC
    -->
    ACE-925M : Not exceed 30A@115V AC or 60A@230V AC, cold start at 25oC
  • Ripple and Noise :
    --> ACE-828M : Less than 50mV@+5V and +3.3V; 100mV@Others
    -->
    ACE-925M : The line regulation for each outputs are less the +/-1%
  • Line Regulation :
    -->
    ACE-828M : Less than +/-2%@+12V, +/-1%@Others
    -->
    ACE-925M : Less than +/-1 for each output
  • Load Regulation :

Model

+5V

+12V

-12V

-5V

+3.3V

+5Vsb

ACE-828M

<+/-3%

<+/-5%

<+/-2%

<+/-2%

<+/-2%

<+/-3%

ACE-925M

<+/-3%

<+/-5%

<+/5%

<+/-5%

 

 

  • Output Power : Total combined output power on the +3.3V and +5V rails is 150W for ACE-828M
  • Efficiency : Higher than 65%
  • Altitude : 0 ~ 10,000ft
  • Protection :
    --> Over voltage protection : The trip point is around 5.7V ~ 7.0V output
    --> Short circuit and overload
  • Vibration : 10 ~ 55Hz at 2G 3min. period, 30min. along X, Y and Z axis
  • Impact : 10G for 20ms once on each X, Y and Z axis
  • Built-in ball bearing cooling fan
  • MTBF :
    -->
    ACE-828M : 177,287 Hours
    -->
    ACE-925M : 132,648 Hours
  • Safety : Meets UL 544, ULl 2601, CSA 22.2 No.601, TUV ( DIN. VDE 075T.1/12.91/EN 60 601-1:1990 )
  • EMI : Meets EN55011 class "B"
  • EMS Standard ( Design to meet the following limits ) :
    --> IEC 801-2 4KV contact, 8KV Air Discharge
    --> IEC 801-3, 3V/M, 26MHz to 1GHz
    --> IEC 801-4, 2KV power mains
    --> IEC 801-5, 2KV criteria B
  • Operating Temperature :
    -->
    ACE-828M : -20o ~ 70oC ( Derating curve will occur 40o ~ 70oC)
    -->
    ACE-925M : 0o ~ 70oC ( Derating curve will occur 40o ~ 70oC)
  • Storage Temperature : -40o ~ 75oC