This image is representative of the product and may not be the exact image of this product.
| Model: ACE-828M |
Manufacturer: IEI | MEDICAL AXT 280W Power Supply |
Product Status: Discontinued - This product is no longer in production. This page is maintained for informational purposes only. Stock Status: Not Available |
|
|
|
|
Input Table :
Model |
Input Range |
Input Frequency |
ACE-828M |
90 ~ 130V AC or 180 ~ 260V AC ( Auto-Switch ) |
47 ~ 63Hz |
ACE-925M |
90 ~ 264V AC ( Full-Range ) |
47 ~ 63Hz |
- Output Table : ( at 40
oC )
Model |
Rated Load |
Max Output |
+5V |
+12V |
-12V |
-5V |
-3.3V |
+5Vsb |
+5V |
+12V |
-12V |
-5V |
-3.3V |
+5Vsb |
ACE-828M |
20A |
10A |
0.5A |
0.5A |
8A |
0.72A |
30A |
15A |
1A |
1A |
15A |
1.2A |
ACE-925M |
25A |
7A |
0.5A |
0.5A |
|
|
30A |
10A |
1A |
1A |
|
|
- Inrush Current :
--> ACE-828M : Not exceed 40A@115V AC or 80A@230V AC, cold start at 25oC
--> ACE-925M : Not exceed 30A@115V AC or 60A@230V AC, cold start at 25oC
- Ripple and Noise :
--> ACE-828M : Less than 50mV@+5V and +3.3V; 100mV@Others
--> ACE-925M : The line regulation for each outputs are less the +/-1%
- Line Regulation :
--> ACE-828M : Less than +/-2%@+12V, +/-1%@Others
--> ACE-925M : Less than +/-1 for each output
- Load Regulation :
Model |
+5V |
+12V |
-12V |
-5V |
+3.3V |
+5Vsb |
ACE-828M |
<+/-3% |
<+/-5% |
<+/-2% |
<+/-2% |
<+/-2% |
<+/-3% |
ACE-925M |
<+/-3% |
<+/-5% |
<+/5% |
<+/-5% |
|
|
- Output Power :
Total combined output power on the +3.3V and +5V rails is 150W for ACE-828M
- Efficiency :
Higher than 65%
- Altitude :
0 ~ 10,000ft
- Protection :
--> Over voltage protection : The trip point is around 5.7V ~ 7.0V output
--> Short circuit and overload
- Vibration :
10 ~ 55Hz at 2G 3min. period, 30min. along X, Y and Z axis
- Impact :
10G for 20ms once on each X, Y and Z axis
- Built-in ball bearing cooling fan
- MTBF :
--> ACE-828M : 177,287 Hours
--> ACE-925M : 132,648 Hours
- Safety :
Meets UL 544, ULl 2601, CSA 22.2 No.601, TUV ( DIN. VDE 075T.1/12.91/EN 60 601-1:1990 )
- EMI :
Meets EN55011 class "B"
- EMS Standard ( Design to meet the following limits ) :
--> IEC 801-2 4KV contact, 8KV Air Discharge
--> IEC 801-3, 3V/M, 26MHz to 1GHz
--> IEC 801-4, 2KV power mains
--> IEC 801-5, 2KV criteria B
- Operating Temperature :
--> ACE-828M : -20o ~ 70oC ( Derating curve will occur 40o ~ 70oC)
--> ACE-925M : 0o ~ 70oC ( Derating curve will occur 40o ~ 70oC)
- Storage Temperature :
-40o ~ 75oC
|
| More Info |
|
|
|
|
|
|
|